EMISSION MEASUREMENT MOBILE LABORATORY

In the design of the system, the selection of the measuring instruments to be installed was based on the widest possible measuring range, a modern and internationally accepted measuring principle, a favourable purchase price and the possibility of computerised data collection. In line with the above, the following instruments were purchased and installed:
 bmelab_25_01.gif

 THC analyzer

Type: MODEL 9000            
Manufacturer: BERNATH ATOMIC
Operating principle: FID
Measuring range: 0-10, 100, 1000, 10000 and 100000 ppm C3H8
Sample volume flow rate: 3 litres/minute

NO/NO2/NOx analyzer

Type: MODEL 42 C HL
Manufacturer: THERMO ENVIRONMENTAL INSTRUMENTS
Operating principle: chemiluminescence
Measuring range: 0-10, 20, 50, 100, 200, 500, 1000, 2000 and 5000 ppm
            0-20, 50, 100, 200, 500, 1000, 2000, 5000 és 10000 mg/m3
Set-up time: 5 sec (NOx measurement)
Sample volume flow: 25 cm3/min

CO analyzer

Type: MODEL 48 C HL
Manufacturer: THERMO ENVIRONMENTAL INSTRUMENTS
Operating principle: IR (Infrared)
Measuring range: 0-50, 100, 200, 500, 1000, 2000, 5000, 10000 and 20000 ppm
            0-50, 100, 200, 500, 1000, 2000, 5000, 10000 és 20000 mg/m3
Setting time: 60 sec.
Sample volume flow: 0,5 - 2 litres/minute

CO2 analyzer

Type: 1400B4 SPX
Manufacturer: SERVOMEX
Operating principle: IR (Infrared)
Measuring range: 1%, 2.5%, 10%, 25%, 50%, two of 100%
Setting time: 30 sec.
Sample volume flow: 1 - 6 litres/minute

O2 analyzer

Type: 1400B4 SPX
Manufacturer: SERVOMEX
Working principle: paramagnetic
Measuring range: two out of 0-5, 10, 20, 25, 50 and 100 V/V %
Settling time: <50 sec.
Sample volume flow: 1 - 6 litres/minute

SO2 analyzer

Type: MODEL 43 C HL
Manufacturer: Thermo Environmental Instruments
Working principle: Fluorescent (PFA)
Measuring range: 0-100, 200, 500, 1000, 2000, 5000 and 10000 ppm
            0-50, 100, 200, 500, 1000, 2000, 5000, 10000 és 20000 mg/m3
Settling time: 80 sec.
Sample volume flow: 1 – 6 liters/min